在控制台中打印HL7消息

时间:2011-11-24 08:19:42

标签: java socket.io hl7

我将一个对象传递给构造函数,然后将该对象的参数添加到HL7。 ORU_R01是HL7的类型。 当我将HL7打印到控制台时,仅打印最后一个OBX。 我的代码出了什么问题? 如何将此HL7消息写入套接字? 在java中有更简单的方法来处理handel HL7吗?

public class FlexSMessageHL7 {
private FileWriter writeHL7ToFile;
private PrismaflexSMessage sMessage;
private ORU_R01 message;
private int i = 0;
private OBX obx = null;

public FlexSMessageHL7(FlexSMessage sMessage) {
    this.sMessage = sMessage;
    this.message = new ORU_R01();
    createHL7SMessage();
}

public void createHL7SMessage() {

    // Populate the MSH Segment
    MSH msh = message.getMSH();
    try {
        msh.getFieldSeparator().setValue("|");
        msh.getEncodingCharacters().setValue("^~\\&");
        msh.getDateTimeOfMessage().setValue(sMessage.getTime().toString());
        msh.getSendingApplication().getNamespaceID().setValue(String.valueOf(sMessage.getMachID()));
    } catch (DataTypeException e) {
        e.printStackTrace();
    }

    // Populate the OBR Segment:time
    OBR obr = message.getPATIENT_RESULT().getORDER_OBSERVATION().getOBR();
    try {
        obr.getObservationDateTime().setValue(String.valueOf(sMessage.getTime()));
    } catch (DataTypeException e) {
        e.printStackTrace();
    }

    // Populate the PID Segment:PatientId
    PID pid = message.getPATIENT_RESULT().getPATIENT().getPID();
    try {
        pid.getPatientID().getIDNumber().setValue(sMessage.getPatID());
    } catch (HL7Exception e) {
        e.printStackTrace();
    }

    // Populate the OBX Segment:Param_Code, time, Measure_Value
    while (i < sMessage.getMsgInfo()) {
        for (PrismaflexSRecord sRecord : sMessage.getsRecordCollection()) {
            try {
                obx = message.getPATIENT_RESULT().getORDER_OBSERVATION().getOBSERVATION(i).getOBX();
                obx.getSetIDOBX().setValue(String.valueOf(i));
                obx.getObservationIdentifier().getIdentifier().setValue(sRecord.getParamCode());
                obx.getDateTimeOfTheObservation().setValue(String.valueOf(sRecord.getTimeStamp()));
                obx.getObservationIdentifier().getNameOfCodingSystem().setValue(String.valueOf(sRecord.getMeasureValue()));
                i++;
            } catch (HL7Exception e) {
                e.printStackTrace();
            }
        }

    }
    try {
        writeHL7ToFile = new FileWriter(File.createTempFile("prismaflexOutputFrom3001HL7", "txt", new File
                ("c:\\tmp\\prismaflex")));
        writeHL7ToFile.write(message.getMSH().toString());
        writeHL7ToFile.flush();
    } catch (IOException e) {
        e.printStackTrace();
    }

    // Now, Encode the message and look at the output
    try {

        Parser parser = new PipeParser();
        String encodedMessage = parser.encode(message);
        System.out.println("Printing HL7 Encoded Message:");
        System.out.println(encodedMessage);
    } catch (HL7Exception e) {
        e.printStackTrace();
    }
}

}

4 个答案:

答案 0 :(得分:1)

您是否考虑过使用HAPI?它是为java编写的,它的对应部分nHAPI也是为.net编写的。详情如下:

http://hl7api.sourceforge.net/

答案 1 :(得分:1)

正如Nicholas Orlowski指出的那样,问题在于行尾字符,根据HL7标准,CR字符使得Windows命令提示只将光标重置为行的开始并用下一行的内容覆盖它。因此,对于控制台输出您需要用其他内容替换行结尾。

对于最近使用HAPI的HL7应用程序,你似乎也在使用它,我做了一个小帮助方法来实现这个功能:

private static String replaceNewlines(String input) {
    return input.replaceAll("\\r", "\n");
}

该函数可以在所有平台上使用,因为它将CR字符替换为特定于操作系统的换行符。

然后我可以用它输出到控制台,如下所示:

LOGGER.trace("Generated message contents:\n" + replaceNewlines(outMessage.encode()));

在这种情况下,我使用log4j登录到控制台,而不是简单的控制台打印输出,但问题对我来说是一样的。

希望它有所帮助!

答案 2 :(得分:0)

我的python HL7py库中遇到过类似的问题。很多时候控制台不喜欢打印字符。我不得不写一个帮助器,将CR更改为LF(换行)以正确显示行。希望有所帮助。

答案 3 :(得分:0)

它不会显示在控制台中,但会在您写入文件时显示。尝试在调试模式下查看变量并将其写入文件。

相关问题