我正在使用PyVISA打开与Agilent电源的通信,当我查询仪器ID时,我不断收到超时错误。
我的代码是:
import visa
rm = visa.ResourceManager()
res = rm.list_resources()
print("Find following resources: ")
print(res)
print("Opening " + res[-1])
inst = rm.open_resource(res[-1])
inst.timeout = 100000 # timeout 100s
inst.write_termination = '\n'
inst.read_termination = '\n'
print(inst.query('*IDN?'))
我得到的输出是:
Find following resources:
(u'ASRL5::INSTR',)
Opening ASRL5::INSTR
Traceback (most recent call last):
File "<ipython-input-10-66ffef23c46a>", line 1, in <module>
runfile('C:/Users/Sabri/Anaconda2/untitled0.py', wdir='C:/Users/Sabri /Anaconda2')
File "C:\Users\Sabri\Anaconda2\lib\site-packages\spyder\utils\site\sitecustomize.py", line 710, in runfile
execfile(filename, namespace)
File "C:\Users\Sabri\Anaconda2\lib\site-packages\spyder\utils\site\sitecustomize.py", line 86, in execfile
exec(compile(scripttext, filename, 'exec'), glob, loc)
File "C:/Users/Sabri/Anaconda2/untitled0.py", line 20, in <module>
print(inst.query('*IDN?'))
File "C:\Users\Sabri\Anaconda2\lib\site-packages\pyvisa\resources\messagebased.py", line 564, in query
return self.read()
File "C:\Users\Sabri\Anaconda2\lib\site-packages\pyvisa\resources\messagebased.py", line 413, in read
message = self._read_raw().decode(enco)
File "C:\Users\Sabri\Anaconda2\lib\site-packages\pyvisa\resources\messagebased.py", line 386, in _read_raw
chunk, status = self.visalib.read(self.session, size)
File "C:\Users\Sabri\Anaconda2\lib\site-packages\pyvisa\ctwrapper\functions.py", line 1584, in read
ret = library.viRead(session, buffer, count, byref(return_count))
File "C:\Users\Sabri\Anaconda2\lib\site-packages\pyvisa\ctwrapper\highlevel.py", line 188, in _return_handler
raise errors.VisaIOError(ret_value)
VisaIOError: VI_ERROR_TMO (-1073807339): Timeout expired before operation completed.
我已经尝试将终止字符指定为读取和写入命令的\ n,指定波特率,将超时更改为“无”,并将仪器声明为inst = rm.open_resource( &#39; ASRL5 :: INSTR&#39;)而不是inst = rm.open_resource(res [-1])。到目前为止,没有任何工作。任何建议将不胜感激!
答案 0 :(得分:0)
尝试使用Keysight(正式的安捷伦)连接专家。
这包含在IO Libraries Suite中: https://www.keysight.com/main/software.jspx?cc=GB&lc=eng&ckey=2175637&nid=-33330.977662&id=2175637
安装完成后,您将能够添加仪器并检查电源的状态。
它是哪种电源型号?
备注:我遇到过PC识别串口COM端口的情况,但只有在我手动更新驱动程序后才能工作。
答案 1 :(得分:0)
我使用Ruby通过RS232端口控制Agilent(keysight)电源e3631a。我相信我有类似的问题。症状是查询命令(如* IDN?)未得到答复,并且错误代码410将显示在设备上。问题是设备无法通过串行端口发送数据,因此无法清除缓冲区。解决方法是在DCB结构中需要设置“ DTR_CONTROL_ENABLE”。在我的情况下,dcb的线路控制部分看起来像“ 0x00000011”。请注意,对于不带流控制的串行通信,通常不需要进行设置。 这个问题使我花了两天时间,希望对其他人有帮助。